Hanna Instruments Checker HC testers are pocket colorimeters: each one measures a single parameter and gives a result shown in digits rather than a colour to match against a card. The table below gives, for each reef parameter, the right Checker, its measuring range, its accuracy according to Hanna and the matching replacement reagent.

Which Hanna Checker for which parameter?

Parameter Checker Range Accuracy at 25 °C Replacement reagents
Alkalinity (dKH) HI772 0.0 to 20.0 dKH ± 0.3 dKH ± 5% of reading HI772-26
Alkalinity (ppm) HI755 0 to 300 ppm CaCO₃ ± 5 ppm ± 5% of reading HI755-26
Calcium HI758 200 to 600 ppm ± 6 ppm ± 5% of reading HI758-26
Magnesium HI783 1,000 to 1,800 ppm ± 5% of reading HI783-25
Phosphate, ultra low range HI774 0.00 to 0.90 ppm ± 0.02 ppm ± 5% of reading HI774-25
Phosphorus, ultra low range HI736 0 to 200 ppb ± 5 ppb ± 5% of reading HI736-25
Phosphate, low range HI713 0.00 to 2.50 ppm ± 0.04 ppm ± 4% of reading HI713-25
Nitrate, high range HI782 0.0 to 75.0 ppm ± 2.0 ppm ± 5% of reading HI782-25
Nitrate, low range HI781 0.00 to 5.00 ppm ± 0.25 ppm ± 2% of reading HI781-25
Nitrite, ultra low range HI764 0 to 200 ppb ± 10 ppb ± 4% of reading HI764-25
Ammonia HI784 0.00 to 2.50 ppm ± 0.05 ppm ± 5% of reading HI784-25
Copper HI702 0.00 to 5.00 ppm ± 0.05 ppm ± 5% of reading HI702-25

Phosphate or phosphorus: HI774, HI736 or HI713?

All three measure orthophosphate; the HI736 expresses it as phosphorus (ppb), the HI774 and HI713 as phosphate (ppm). The two readings cannot be compared directly.

  • The HI774 shows phosphate directly in ppm, from 0.00 to 0.90 ppm. Hanna designs it for marine systems, including reef tanks.
  • The HI736 shows phosphorus in ppb. Hanna recommends it for SPS coral tanks and aquariums aiming for phosphate below 0.03 ppm. To convert, Hanna gives the formula: ppb phosphorus × 3.066 ÷ 1,000 = ppm phosphate.
  • The HI713 covers 0.00 to 2.50 ppm. Hanna lists it for fish-only tanks with live rock and reefs with corals less sensitive to phosphate, and notes that its accuracy can be problematic for certain reef aquarists.

Alkalinity: in dKH or in ppm?

The HI772 and the HI755 use the same method; only the displayed unit changes. According to Hanna, 1 dKH = 17.86 ppm CaCO₃. Choose the unit you already dose in.

Nitrate: low or high range?

The HI782 reads directly from 0.0 to 75.0 ppm. The HI781 reads from 0.00 to 5.00 ppm with a resolution of 0.01 ppm; above that, Hanna provides a 1:10 dilution procedure covering up to 50.0 ppm. Both were developed to work with seawater.

Salinity: the HI98319 tester

The Hanna HI98319 shows salinity in ppt, PSU or specific gravity (S.G.), with automatic temperature compensation. Its accuracy is ± 1 ppt between 0 and 40 ppt. It calibrates at a single point, 35 ppt, with the Hanna solution: box of 25 HI70024P sachets or a single sachet.

Frequently asked questions

Does a Hanna Checker need calibrating?

No. You zero it with the water sample before each measurement, then add the reagent. Hanna offers standards to check the meter, not to calibrate it.

Do Hanna Checkers work in saltwater?

The models on this page are the ones Hanna designs for saltwater. The HI713 and HI702 are listed for fresh or saltwater.

What exactly does the HI784 ammonia Checker measure?

Total ammonia (NH₃ and NH₄⁺). Hanna's manual includes a table, by pH and temperature, to work out the share of toxic NH₃ ammonia.

What battery and what warranty?

Checker HC testers run on one AAA battery and carry a 1-year Hanna warranty.

Can the cuvettes be replaced?

Yes: the HI731321 16 mm glass cuvettes fit Checker HC testers.